Seaward RIGEL SP-SiM SPO2 Simulator with PULS-R Finger Simulator, its actual photo displayed.
- Brand NEW still in original box.
- Made in 2018
- Supplied with standard accessories.
- Can be Offered with a fresh calibration certificate at extra cost.
- Features:
- The Rigel SP-SIM is a hand held, high performance pulse oximetry simulator for the fast, accurate and efficient testing of SpO2 devices. Utilising both optical and electronic simulation techniques, the SP-SIM SpO2 simulator eliminates the inconsistencies that can often be associated with testing using traditional equipment like an optical finger.
- With full adjustments of pulse volume and heart rate, the Rigel SP-SIM SpO2 simulator is compatible with most common R-curves for maximum flexibility.
- Bluetooth connectivity to a range of accessories and Med-eBase, comprehensive database software, ensures fast and easy download of test results, managing your asset database, creation of test sequences and producing test certificates.
- PULS-R SpO2 Finger Simulator
-
- The compact Rigel PULS-R universal SpO2 simulation finger is the latest accessory to the Rigel UNI-SIM and SP-SIM vital-signs simulators.
- Able to simulate accurate SpO2 simulations in 1% resolution from as low as 30% using the pre-programmed manufacturer specific R-curves, the PULS-R can also be configured to meet customer specific R-curves using the simple to use configuration tool.
- Supported default R-curves:
-
|
|
Click here to view a range of accessories available for the Rigel SiM range
RIGEL SP-SiM Technical Specification:
SpO2 Simulation | |
---|---|
Simulation | Optronic (Electronic & Optical), indirect via probe adaptor box |
Simulation via | Probe & full chain |
Range | 50 to 100% |
Accuracy | ± 0.5% of reading between 80 – 100% SpO2 ± 1% of reading between 50 – 79% SpO2 |
Heart Rate | 20-300 bpm |
Accuracy | ± 1 bpm |
R-curves | Datex, Datex Ohmeda, Nellcor, N.Oximax, Masimo,Criticare, Datascope, Philips (HP), CSI, Novametrix, Nihon Kohden, Nonin, SensorMedics (upgradeable) |
Perfusion Index | 0-20% (5% default) |
Arrhythmias | Tachycardia, Bradycardia, (upgrade) |
Artefacts | Motion, light (AC-DC) (upgrade) |
Direct Interface | To Probe adaptor box |
Chronometer | Test option to test response time SpO2 monitor |
General Specification | |
---|---|
Operation | Battery cell, insitu charge |
Mains Supply | 230V/110V AC +/- 10% 50/60Hz |
Battery Life | >10 hours testing |
Weight | <1 kilo |
Dimensions | 26 x 10 x 5.5 cm |
Operating Conditions | 10-30’C, 0-90% RH – NC |
Storage Environment | -15′ – +60’C |
Environmental Protection | IP 40 |
Part Number: 370A920 | |
---|---|
PULS-R Technical Specification:
Accuracy of simulation when used with the corresponding R-curves
Resolution | Range | Repeatability* |
1% steps | 30-60% | ±5%** |
1% steps | 60-90% | ±3% |
1% steps | 90-100% | ±1% |
*Based on using the same probe and monitor setup
**Note that some monitor types might not be able to display low range sats
Operating temperature | 0-40°C |
Dimensions | 78 x 68 x 29 mm |
Finger shape | 40 x 18 mm |
Weight | 200 grams |
Part number: 399A910