The IVY Biomedical 7800 Cardiac Trigger Monitor is a specialised, high-precision cardiac monitoring device designed to provide reliable R-wave detection and accurate cardiac triggering for synchronised medical procedures. Commonly used in MRI, CT, and nuclear medicine environments, the 7800 ensures stable cardiac gating signals to support high-quality imaging and treatment workflows.
Engineered for dependable performance, the unit features advanced ECG signal filtering, adjustable trigger thresholds, and real-time waveform display, allowing clinicians to achieve optimal synchronisation even in challenging monitoring conditions. Its compact, lightweight design makes it easy to integrate into any clinical setup.
Technical Specifications:
| Specification | Details |
|---|---|
| ECG Configuration | 4-lead system |
| Trigger Lead Options | Lead I, II, III, or AUTO (menu selectable) |
| Second Lead Display | Lead I, II, or III |
| Patient Cable | 4-lead cable with 6-pin AAMI connector |
| Patient Isolation | > 4 kV rms, 5.5 kV peak |
| Frequency Response (ECG) | 0.67 – 100 Hz (unfiltered) |
| Filtered Response | 1.5 – 40 Hz; auto 50/60 Hz notch filter |
| CMRR | ≥ 90 dB |
| Tall T-Wave Rejection | ≤ 1.2 × R-wave amplitude |
| Pacer Pulse Rejection | 0.1 – 2 ms width @ ±2 to ±700 mV (user selectable) |
| Defibrillator Protection | 360 J discharge, < 5 sec recovery (Type CF) |
| Heart Rate Range | Adult: 10–300 bpm; Pediatric/Neonate: 10–350 bpm |
| Heart Rate Accuracy | ±1% ± 1 bpm |
| HR Resolution | 1 bpm |
| Sensitivity | 300 µV (peak) |
| HR Averaging | Exponential @ 1 Hz; 2 or 8 s response time |
| Alarms | High Rate: 15–250 bpm (5-bpm steps) Low Rate: 10–245 bpm (5-bpm steps) Asystole: R-R > 6 sec Lead Off / Check Lead |
| ECG Simulator | Built-in; 1 mV amplitude, 10–250 bpm |
| Display | 8.4″ TFT colour touch-screen LCD, 640×480 resolution |
| Synchronized Output | < 2 ms delay; R-R accuracy ±75 µs typical; Output impedance < 100 Ω |
| Trigger Output Settings | Pulse width: 1 / 50 / 100 / 150 ms Amplitude: 0–+5 V or –10 to +10 V Polarity: Positive or Negative |
| Interfaces / I/O | BNC Sync Output ¼″ ECG Analog Output DB-9 X-ray/Aux RS-232 Ethernet (10Base-T) USB Memory Stick |
| Optional Recorder | 2-trace direct-thermal strip chart recorder |
| Dimensions (H × W × D) | 22.14 × 23.50 × 15.49 cm (8.72″ × 9.25″ × 6.10″) |
| Weight | 2.54 kg (5.6 lb) |
| Power Requirements | 100–120 V or 200–230 V AC, 50/60 Hz; 45 VA max |
| Operating Conditions | 5°C–40°C; 0–90% RH (non-condensing); altitude –100 to +3,600 m; IPX1 |
| Storage Conditions | –40°C to +70°C; 5–95% RH; altitude –100 to +14,000 m |
| Regulatory Standards | IEC 60601-1, IEC 60601-2-27, AAMI EC-13, CE, ISO 13485 |
1X Used unit in stock. £990 for unit alone without stand.
Immediately available for dispatch, patient ready and this product is offered with a 6 month warranty period.
Graseby 3000 Series Case Kits -New
Alaris SP Chassis Enhancement Kit MKII 1000SP01136
Argus MCM 404 Infusion pump
New Carefusion IVAC® PCAM® Syringe Pump Upgrade Cover Shield & Stickers Kit REF: 1000SP01346
Codan Drop Detectors - New
Graseby Omnifuse PCA Syringe Pumps
HP M1700-69502 IEC ECG Patient Module 