RIGEL SP-SiM SPO2 & PULS-R Finger Simulators – New

Availability:

1 in stock


£1,780.00

1 in stock

Seaward RIGEL SP-SiM SPO2 Simulator with PULS-R Finger Simulator, its actual photo displayed.

  • Brand NEW with warranty.
  • Offered with a fresh calibration certificate.
  • Made in 2018
  • Supplied in its original box with standard accessories.

The Rigel SP-SIM is a hand held, high performance pulse oximetry simulator for the fast, accurate and efficient testing of SpO2 devices. Utilising both optical and electronic simulation techniques, the SP-SIM SpO2 simulator eliminates the inconsistencies that can often be associated with testing using traditional equipment like an optical finger.

With full adjustments of pulse volume and heart rate, the Rigel SP-SIM SpO2 simulator is compatible with most common R-curves for maximum flexibility.

Bluetooth connectivity to a range of accessories and Med-eBase, comprehensive database software, ensures fast and easy download of test results, managing your asset database, creation of test sequences and producing test certificates.

PULS-R SpO2 Finger Simulator

The compact Rigel PULS-R universal SpO2 simulation finger is the latest accessory to the Rigel UNI-SIM and SP-SIM vital-signs simulators.

Able to simulate accurate SpO2 simulations in 1% resolution from as low as 30% using the pre-programmed manufacturer specific R-curves, the PULS-R can also be configured to meet customer specific R-curves using the simple to use configuration tool.

Supported default R-curves:

  • Beijing Choice
  • Criticare
  • GE Tuffsat
  • Masimo
  • Mindray
  • Nellcor
  • Nellcor Oximax
  • Nihon Kohden
  • Nonin
  • Novametrix
  • Phillips / HP

Click here to view a range of accessories available for the Rigel SiM range

RIGEL SP-SiM Technical Specification:

SpO2 Simulation
Simulation Optronic (Electronic & Optical), indirect via probe adaptor box
Simulation via Probe & full chain
Range 50 to 100%
Accuracy ± 0.5% of reading between 80 – 100% SpO2
± 1% of reading between 50 – 79% SpO2
Heart Rate 20-300 bpm
Accuracy ± 1 bpm
R-curves Datex, Datex Ohmeda, Nellcor, N.Oximax, Masimo,Criticare, Datascope, Philips (HP), CSI, Novametrix, Nihon Kohden, Nonin, SensorMedics (upgradeable)
Perfusion Index 0-20% (5% default)
Arrhythmias Tachycardia, Bradycardia, (upgrade)
Artefacts Motion, light (AC-DC) (upgrade)
Direct Interface To Probe adaptor box
Chronometer Test option to test response time SpO2 monitor
General Specification
Operation Battery cell, insitu charge
Mains Supply 230V/110V AC +/- 10% 50/60Hz
Battery Life >10 hours testing
Weight <1 kilo
Dimensions 26 x 10 x 5.5 cm
Operating Conditions 10-30’C, 0-90% RH – NC
Storage Environment -15′ – +60’C
Environmental Protection IP 40
Part Number: 370A920

PULS-R Technical Specification:

 Accuracy of simulation when used with the corresponding R-curves

Resolution Range Repeatability*
1% steps 30-60% ±5%**
1% steps 60-90% ±3%
1% steps 90-100% ±1%

*Based on using the same probe and monitor setup
**Note that some monitor types might not be able to display low range sats

Operating temperature 0-40°C
Dimensions 78 x 68 x 29 mm
Finger shape 40 x 18 mm
Weight 200 grams

Part number: 399A910