This is a specific, low-cost remedial kit issued by Caesarea Medical Electronics (CME) Ltd. to service the NIKI T34 Ambulatory Syringe Pump. The kit contains adhesive foam pads and battery insertion labels/stickers. It is used to perform a mandatory field update to improve the reliability of the pump’s power supply by preventing the 9V battery from shifting. The labels, which prominently instruct the user to “insert battery end against foam pad”, ensure that the battery is correctly and securely seated against the contacts after the modification is made.
Technical Specifications:
| Parameter / Feature | Specification |
| Manufacturer | Caesarea Medical Electronics (CME) Ltd. |
| Compatible Device | CME NIKI T34 and T34 Ambulatory Syringe Pumps |
| Reference (REF) Number | OKT00010 |
| Component Type | Remedial Kit (Adhesive parts for battery compartment) |
| Contents | Adhesive Foam Pads and Battery Insertion Label(s)/Stickers |
| Function/Purpose | This kit provides the necessary materials for a mandatory service update to prevent loss of battery connection due to varying battery lengths. The foam pads stabilize the 9V battery against the contacts. |
| Label Instruction | The included sticker/label provides the critical instruction to the user: “insert battery end against foam pad”, ensuring correct battery placement after the foam pad is installed. |