Fluke QA-ES II Electrosurgical Analyzer, in very good operational and physical condition, analyser’s actual photos displayed.
- Made in USA
- 115-140VAC 50/60Hz
- Revision: 2.26
- Fresh calibration certificate can be offered at extra cost.
- Supplied with: Operating Manual
- FEATURES:
- QA-ES Series II analyzes electrosurgical units quickly and accurately.
- Automatic power distribution measurement, including: power current, Peak-to-peak voltage (closed load only) and Crest-factor
- Oscilloscope output
- High-frequency leakage measurements with an accuracy of ±5 % of reading down to 20 mA guarantees reliable high-frequency leakage results.
- A wide load-resistance range provides 128 user-selectable loads from 10 Ohm to 5200 Ohm, including very low loads for testing many of today’s ESUs.
- Foot-switch output for triggering the ESU under test
- An Ansur QA-ES software plug-in allows users to create and automatically run tests, capture data, and produce easy-to-read reports with a PC.
- Large display
- RS-232 and Centronic-printer interface
- Technical specifications:
- Generator output:
- Continuous operation: Continuous measurement of power, current, peak-to-peak voltage (closed load only), and crest factor
- Single operation: Single measurement after the set delay time of the ESU output of power, current, peak-to-peak voltage (closed load only), and crest factor
- Power distribution: Automatic measurement of power, current, peak-to-peak voltage (closed load only), and crest factor through a user-selectable load range
- RF leakage current: Provides connections and load configurations to measure HF leakage from both grounded and isolated equipment
- RECQM: Test the “return electrode control quality monitoring” using the QA-ES internal loads
- Modes of operation: Manual or remote controlled (via Ansur)
- Measurement: True-rms value of applied waveform
- RMS bandwidth
- 30 Hz to 10 MHz (-3 dB) for instrumentation only
- 30 Hz to 2.5 MHz (-3 dB) with loads
- Low frequency filter: 100 Hz filter to avoid low-frequency disturbance or interference
- Current: 20 mA to 2200 mA
- Current accuracy: 20 mA to 2200 mA ± 2 % of reading
- Load resistance
- 10 Ω to 2500 Ω in step of 25 Ω (@ dc)
- 2500 Ω to 5200 Ω in step of 100 Ω (@ dc)
- Additional fixed load: 200 Ω 400 W for 30 s; max 15 % duty cycle
- Crest factor: The higher of the two peak voltage measurements is used for computation
- Range: 1.4 to 16 (V peak/V rms)
- Foot-switch output: The foot switch output can be used to trigger the electrosurgical unit
- Peak-to-peak voltage
- 0 kV to 10 kV (closed load only)
- accuracy: ± 10 %
- Note: Measurement is taken between the active and dispersive electrodes with closed load only.
- Oscilloscope output: 5 V/A uncalibrated, 100 mA RF current minimum input
- Ansur QA-ES Plug-In:
- User-programmable test sequences: Allows unlimited numbers of test sequences with user-programmable templates and test limits. These tests include power distribution test, output test, HF leakage, and RECQM verification
- Remote control: All functions and tests in QA-ES may be performed from the PC
- Storage and recall: Protocol formats and data may be stored, recalled, printed out, or transferred
- Storage: 0 °C to 50 °C (32 °F to 122 °F)
- Temperature Operating: 15 °C to 35 °C (59 °F to 95 °F)
- Display: LCD graphic display
- Alphanumeric format: 8 lines x 40 characters
- Graphic mode: 240 x 64 pixel matrix
- Display control: 5 F-keys, enter, cancel, control knob, and an encoder
- Data input/outputs: Parallel printer port and bidirectional RS-232
- Power: 115/230 V ac; 48 Hz to 66 Hz, 35 VA
- Housing: Metal case
- Dimensions (W x D x H): 39.5 cm x 34.2 cm x 13.2 cm (15.6 in x 13.5 in x 5.2 in)
- Weight: 9.8 kg (21.6 lb
- Generator output: