Rigel 288 Electrical Safety Analyser

Availability:

Out of stock


SOLD OUT

Seaward Rigel 288 hand held electrical safety tester, in good condition, similar to photos attached.

  • Supplied with:
      • Patient Applied part module
      • 10 Applied part adaptors
      • Earth bond test probe with clip
      • Detachable 2 meter mains cable
      • Carry case
      • Operating manual with software utilities. 
      • Other accessories can be added at extra cost, email us for prices.
      • Fresh calibration can be provided at extra cost.
  • Features: 
    • Light, Hand-held With its purpose designed robust enclosure, the Rigel 288 is truly handheld. It is easy to hold single-handedly and enables one hand operation and navigation, mains and battery operation.
    • Versatile: Test in Accordance with the leakage requirements of IEC/EN 62353, IEC/EN 60601-1, AAMI, NFPA, AS/NZS 3551 and VDE 0751-1 using separate IEC 60601 and AAMI Body Model
    • Easy to use: A full graphic, monochrome LCD display (1/4 VGA minimum) in combination with an integral alpha-numeric ABCD-key-board.
    • Full Automatic test and Manual test modes: Able to perform UTS (Unique Test Sequence) and allows fully automatic, semi automatic and fully manual testing.
    • User definable test routines: Users have the ability to amend the default programs or create new programs by copying the preset test programs. Each program will have a unique Identifier
    • Multiple Applied Part Function: This feature gives the user the capability of testing up to 10 individual Applied Parts from different Modules or classes e.g. BF and CF class, or Bf ECG and Bf SPO2 module.
    • Test report generated can be sent from Rigel 288 to mobile phone or PC via Bluetooth or RS 232
    • Windows Explorer type user interface – layout, report can be displayed and printer in PDF format, Output of database to Excel / Access.
    • Test schedule function
    • Printing of test certificate
    • Store test certificate as HTML for easy email application
    • Internal Asset management facilities: Store up to 10,000 test records, custom test routines, visual inspections and performance tests and download to and from PC via Bluetooth Interface.
  • Applications:
    • Routine testing of Medical Electrical Equipment
    • Service tool for Performance Testing
    • Asset Management
    • Fast and efficient testing of IEC leads
    • Earth bond testing on (Medical) Installations and non Medical Equipment
    • Electrical Safety Tests performed:
      • Earth bond*
      • Insulation
    • Specific to IEC 60601-1:
      • Earth Leakage
      • Enclosure Leakage
      • Patient Leakage
      • Patient Auxiliary
      • Patient F-type
    • Specific to pr IEC 62353 – VDE 0751:
      • Equipment Leakage (Direct, Differential and Alternative Method)
      • Applied Part Leakage (Direct and Alternative Method)
    • Specific to AAMI & NFPA:
      • Patient Leakage AP-GND
      • Patient Leakage AP-CASE
      • Patient Leakage AP- AP
      • Patient Leakage AP- ALL (AAMI / NFPA)
    • Custom tests can be created using a variation or combination of the above.
  • SPECIFICATIONS:
    • Earth Continuity:
      • Method: 2 wire technique, using ‘zero’ lead function.
      • Test Current: >+200mA, -200mA DC into 2 ohms
      • Max Test Voltage: 4-24V rms o/c (6V for IEC 60601 tests)
      • Measuring Range (low range): 0.001 – 0.999 ohms @ 0.001 ohms resolution
      • Measuring Range (mid range): 1.00 – 9.99 ohms /@ 0.01 ohms resolution
      • Measuring Range (high range): 10.0 – 19.9 ohms @ 0.1 ohms resolution
      • Accuracy: ± 3% of reading + 10 m ohms
    • Insulation Resistance:
      • Measurement EUT to Earth / Ground, EUT to AP, AP to Ground
      • Voltage: 250V DC, 500 V DC @1mA.
      • Range (low range): 0.01Mohms – 20 Mohms
      • Accuracy (low range): ± 5% of reading +2 counts
      • Range (high range): 20Mohms – 100Mohms
      • Accuracy (high range): ±10% +2 counts
      • Resolution: 0.01Mohms
    • Direct Leakage Measurement
      • Measuring Range: 4µA to 9999µA
      • Accuracy: ± 5% or reading +2 counts
      • Mains on A.P. voltage: F-type only @ 110% of mains
      • Measuring Device: As per IEC 60601-1 requirements
      • Measurement Type: Separate AC & DC for Patient (-Auxiliary) True RMS for all remaining Leakage tests Leakage to IEC 60601
    • Differential Leakage Measurement
      • Measuring Range: 75µA to 9999µA
      • Accuracy: ±5% of reading + 5 counts
      • Measurement / display resolution: 1µA
      • Measurement Type: True RMS
      • Measuring Device: Similar frequency response characteristics to IEC 60601-1.
    • Alternative Leakage Measurements
      • Test Voltage: 250V at mains frequency
      • Test Current: 3.5 mA current limited
      • Measurement Range: 4µA to 9999µA
      • Measurement Resolution: 1µA
      • Measurement Accuracy: ±5% of reading + 2 counts
      • Measurement Type; True RMS
      • Measuring Device: As per IEC 60601-1
    • Power Measurement
      • Method: VA rating.
      • Range: 0.1KVA – 4KVA
      • Accuracy: ±10% + 2 counts
    • Mains Outlet Test
      • Input voltage range: 0-300V AC, max current 16A
      • Measures: L – E, N – E & L – N
      • Accuracy: ± 5% of reading + 2 counts
    • IEC Mains Lead Test
      • Test Duration: 2s
      • Test: Continuity of all conductors, Earth bond, Insulation & Polarity
    • General
      • Mains power: 230 VAC ±10%, 50Hz +/- 1Hz 120 VAC ±10%,
      • Battery 6 x 1.5V Alkaline AA
      • Weights: 1.6 kg including batteries
      • Size (L x W x D): 270 x 110 x 75 mm / 10.5 x 4 x 3”
      • Operating conditions: 0′ – 40’C, 0-90% RH – NC
      • Storage environment: -15′ – +60’C
      • Environmental Protection: IP 40

 

error: