Diomed Delta 15 Laser – Diode

Availability:

Out of stock


SOLD OUT

AngioDynamics Diomed DELTA 15 Diode Laser unit in very good physical and operational condition, unit’s actual photo displayed.

  • Serial No. 003596
  • Made In: UK
  • Date of manufacture 12/2006
  • Voltage 100-240V, 500VA, 50/60Hz
  • Version: 41.16

*** Supplied with:

  • Diomed delta Foot switch
  • User manual
  • 1 x key
  • 1 x remote interlock bypass

*** Key features of the DIOMED DELTA 15 include:

  • • 15W power output
  • • Compact & portable
  • • Intuitive user interface
  • • Automatic procedure recognition when used with fibers and accessories conforming to the
    DIOMED FRS fiber system
  • • Minimal maintenance & service
  • • DELTA 15 has been designed for use with the DIOMED range of procedure kits, such
    as those for EVLT (EndoVenous Laser Treatment). It can also be used with a wide range of standard fibers and accessories.

*** Indications: The DIOMED DELTA 15/30 is intended for the following contact or non-contact
laser procedures:

  • • Incision (only contact)
  • • Excision (only contact)
  • • Vaporisation
  • • Coagulation / haemostasis
    – of soft tissue in Endovascular, open and endoscopic procedures in Vascular
    Surgery, General Surgery, Gynaecology, Urology, Otorhinolaryngology
    (ENT)/Head & Neck Surgery, Ophthalmology (oculoplastics),
    Pulmonology/Thoracic surgery, Plastic Surgery, Gastroenterology and
    Neurosurgery.

*** SPECIFICATIONS: 

  • Laser Type: Diode laser, CW
  • Centre wavelength: 810 nm
  • Spectral bandwidth: 90% of optical power contained within 810 ± 20 nm
  • Delivery fiber: Minimum 400 µm diameter, 0.37 Numerical Aperture (NA)
  • Fiber connector: Optical SMA-905 to MIL STD
  • Maximum Power: 15 W to the laser output port
  • Minimum Power: 0.5 W to the laser output port
  • Set power accuracy Better than ± 10% of displayed power
  • Output power stability: ± 5% maximum
  • Aiming beam: Red Class 3a diode laser, ≤ 5 mW at port, wavelength 635-655 nm
  • Pulse duration: 100 ms – 9900 ms, Continuous Wave (CW)
  • Pulse Interval: 100 ms – 1000 ms
  • Countdown timer: 10 – 3200 seconds
  • Fibers: Contact fibers, 400µm, 600µm and 1000µm, Non-contact fibers, 600µm & 1000µm
  • Calibration: Internal power calibration meter for handpiece or fiber calibration (Optional on DIOMED DELTA 15)
  • Spot Handpiece Mode: Spot handpiece mode is available on the DIOMED DELTA 15 when fitted with the optional calibration meter
  • Spot size: Ø 2mm 
  • Single/Repeat Pulse: 50ms – 250ms 
  • Pulse Interval: 100ms – 1000ms
  • Fluence Up to: 119J/cm2
  • Fiber Recognition System: (FRS)
  • Operating Frequency: 125 kHz ± 1 kHz
  • Operating Range: Maximum of 40 mm from the laser output port
  • RFID Modulation Type: A1D
  • Standards: FCC 47 Part 15c, Canadian RSS-210 EN 300 330-1, EN 300 330-2, EN 301 489-1, EN 301 489-3
    Cooling By ambient air, with fan assistance
  • Power Supply: 100 – 240 V AC, 50 – 60 Hz, 500 VA max.
  • Dimensions (H x W x D): 150 mm x 370 mm x 325 mm (± 5 mm)
  • Weight: 12 kg max.
  • Operating Temperature: +10ºC to +40ºC
  • Operating Humidity: Up to 75% relative humidity, non-condensing
  • Operating Pressure: Normal atmospheric pressure
  • Storage Temperature: 0ºC to +55ºC
  • Storage Humidity: Up to 90% relative humidity, including condensing
  • Safety Standards: EN 60601-1, EN 60601-1-2, EN 60601-2-22, EN 60825-1, 21 CFR 1040.10, 1040.11
Category: